Invention Disclosure Review
Balloon-Expandable Intraluminal Stent
- Generated
- Jul 20, 2026
Generated on a public invention disclosure — no confidential material involved. First-pass analysis — not an opinion of counsel or drafted claims.
Cross-Domain Application Map
Endovascular / interventional cardiology (origin field) — Balloon-expandable scaffold for maintaining patency of coronary, iliac, femoral, renal and other blood vessels after angioplasty, with final diameter set operator-side against each vessel wall.
This is the direct embodiment of the generalized function: a compact tube delivered into a lumen and plastically expanded beyond the elastic limit so it resists chronic vessel recoil without spring-back, with the wall-set diameter solving the migration/over-expansion failure of fixed-diameter self-expanding grafts. Highest commercial pull and broadest in-field breadth, but this is also where the closest and most threatening prior art (self-expanding grafts, coil/spring endoprostheses, balloon angioplasty) lives, for counsel to assess.
Evidence: described_not_demonstrated · Application distance: near
Oil & gas well construction / downhole completions — Expandable tubulars, liners, and sand screens run compact into an irregular open borehole and plastically expanded (by mandrel/cone or internal pressure) to line, seal, and structurally support the wellbore wall at a diameter dictated by the hole.
Identical mechanism in a remote field: the borehole is a non-uniform 'lumen' whose final supported diameter cannot be predetermined at manufacture, so in-situ plastic expansion to conform to and prop the wall — beyond the elastic limit so there is no recovery under formation load — is exactly the value proposition. Very high commercial pull and far from cardiology, making it a strong non-obvious moat piece; the structural-geometry trade-off (perforated/slotted vs. mesh, radial strength vs. deliverability) maps directly onto screen vs. solid-liner choices.
Evidence: speculative · Application distance: far
Non-vascular medical lumens (GI, hepatobiliary, urology, pulmonology) — Balloon-expandable scaffolds for biliary ducts, ureters/urethra, esophagus, colon, and tracheobronchial airways where a stricture must be held open at a lumen-matched diameter.
The disclosure expressly contemplates 'other body passageways'; each is a soft-walled lumen with patient-to-patient variation, so operator-controlled in-situ plastic expansion that conforms to the individual wall and resists collapse delivers the same benefit as in vessels. Broad medical breadth and real clinical pull; low delivery-profile constraint and metal-to-wall coverage trade-off carry over directly.
Evidence: described_not_demonstrated · Application distance: near
Downhole remediation / pipeline integrity (subsurface) — Expandable metal patch or sleeve deployed compact across a corroded, split, or leaking casing/tubing section and plastically expanded to seal the defect and restore load-bearing wall support (zonal isolation, casing patch).
A repair-driven variant of the same function: the host bore's damaged inner wall sets the achievable expanded diameter, and permanent plastic deformation is essential so the patch stays apposed and pressure-sealed after the expansion tool is withdrawn — the exact 'no spring-back, wall-set diameter' requirement. Distinct commercial driver (remediation vs. new completion) makes it a separately-filable/licensable target with high field-of-use value.
Evidence: speculative · Application distance: far
Mining / tunneling geotechnical support (rock reinforcement) — Expandable friction rock bolt / ground anchor: a folded thin-walled metal tube inserted into a drilled borehole and inflated to plastically conform to the irregular rock wall along its length, providing frictional anchoring at a hole-set diameter.
Maps the function one-for-one — compact tube into a bore, internal dilating force, plastic deformation beyond the elastic limit, permanent conformance to a wall whose profile is not known in advance. Strong functional fit and real commercial pull in mining/civil tunneling; note that an established commercial analog of this expansion-in-borehole principle exists in this field, which is directly relevant to counsel's prior-art assessment of the raw function.
Evidence: speculative · Application distance: far
Civil water/wastewater infrastructure (trenchless rehabilitation) — Internal expandable metal liner or sleeve pulled into a corroded/leaking buried pipe in a reduced-diameter state and plastically expanded to line and structurally reinforce the host pipe at its actual (variable) internal diameter.
The aging pipe is a lumen of unknown, non-uniform ID; permanent in-situ expansion that conforms to and props the host wall — rather than a factory-fixed liner size — solves fit and standoff problems the same way the stent solves vessel apposition. Large addressable infrastructure market and far from the origin field.
Evidence: speculative · Application distance: far
Claim Architecture
The claim-level view (the primary analysis): the broadest defensible claim is on the mechanism itself — it covers every use — while a standalone same-mechanism application claim is weak. This is engineering work-product / a scaffold for counsel, NOT drafted claims or a patentability conclusion. Grounded in the prior art available to this report; confirm against a full search before relying on it.
Broadest mechanism (the dominating claim): A radially deformable tubular metal scaffold that is delivered into a body lumen in a small crimped state on an internal dilating member and then plastically deformed in situ — beyond the elastic limit of its constituent metal — by an internally applied radial force, so that it permanently retains an expanded diameter that is set by the surrounding lumen wall rather than by a shape predetermined at manufacture. The mechanism is defined by (a) a plastically (not elastically) deformable open tubular metal body, (b) crimping to a delivery diameter over an internal balloon/dilator, and (c) permanent retention of an operator-selected, wall-limited expanded diameter after the dilating force is removed.
Structure claim set
Covers: tubular prosthesis formed as a wire-mesh tube, tubular prosthesis formed as a slotted thin-walled tube, radial plastic deformation with force in excess of the elastic limit of the material, tube retains expanded diameter d' after balloon deflation and catheter withdrawal, final implanted diameter d' set in situ by the vessel wall, not predetermined at manufacture
Closest prior art: JP2010246987A — an unexpanded stent whose tubular wall of interconnected struts undergoes plastic deformation to a maximum allowable point on application of a radially outward force (a balloon-expandable, plastically deformable metal tube). This is the retrieved reference that most threatens the core plastic-deformation structural concept. (Attorney to verify publication/priority date against the disclosure — this and the other retrieved patents appear to post-date the foundational concept.) (verify)
Distinguishing limitation: Do NOT rest the distinction on 'plastic deformation to a retained diameter' alone — JP2010246987A already discloses a plastically deformable strut/slotted tube. The limitation that survives the closest reference is the wire-mesh tubular construction as the deformable body (an interlaced/mesh wall rather than a strut-defined slotted or cut-tube wall), together with the expanded diameter being limited by the lumen wall rather than by a device-defined 'maximum allowable point.' The slotted-tube embodiment standing alone is closely approached by JP2010246987A and JP5908937B2 and should be claimed with narrower geometry (coverage-ratio / cell configuration) rather than as a bare slotted tube.
Standard: §102 anticipation / §103 obviousness (for counsel to assess)
Composition claim set
Covers: biocompatible metal candidates: implant-grade stainless steel, tantalum, silver, gold, titanium, radial plastic deformation with force in excess of the elastic limit of the material
Closest prior art: US6059810A — balloon-mounted, mechanically expanded stent, but composed of a shape-memory (Nitinol/martensitic-austenitic) alloy that expands toward a set austenitic diameter. (verify)
Distinguishing limitation: The prosthesis body is a non-shape-memory, work-hardening implant-grade metal (e.g., stainless steel, tantalum, titanium) that holds its expanded diameter purely by plastic deformation, with no temperature- or phase-driven recovery diameter — in contrast to US6059810A, whose recovery/expanded diameter is materially predetermined by the alloy's austenitic set shape. Note for counsel: a bare recitation of common biocompatible metals is weak on its own; the durable content is the pairing of these metals with the wall-limited plastic-retention mechanism, not the material list.
Standard: §103 obviousness (for counsel to assess)
Method-of-use claim set
Covers: tube crimped over angioplasty balloon catheter at a small first (delivery) diameter d, radial plastic deformation with force in excess of the elastic limit of the material, tube retains expanded diameter d' after balloon deflation and catheter withdrawal, final implanted diameter d' set in situ by the vessel wall, not predetermined at manufacture, operator-controlled, variable expansion ratio d'/d matched to individual body passageway
Closest prior art: D3 (balloon angioplasty catheter dilation of stenotic vessels) combined/contrasted with D1 and D2 (self-expanding grafts and coil/spring endoprostheses that expand to a single predetermined diameter). Among retrieved art, JP2010246987A is the nearest balloon-expandable deployment reference. (verify)
Distinguishing limitation: The step of plastically deforming the crimped metal tube in situ to an operator-selected diameter that is limited/defined by the surrounding lumen wall (variable expansion ratio d'/d chosen at the point of use), such that a single delivered device permanently conforms to a range of lumen sizes — as opposed to D1/D2, which self-expand to one factory-preset diameter, and D3, which dilates the vessel without leaving a permanently deformed supporting scaffold. Frame the distinction on the wall-limited, operator-set final diameter and the permanent plastic retention after withdrawal, not on 'balloon delivery' generally (which D3 and JP2010246987A disclose).
Standard: §103 obviousness (for counsel to assess)
Process claim set
Covers: tubular prosthesis formed as a wire-mesh tube, tubular prosthesis formed as a slotted thin-walled tube, structural geometry trade-off: wire mesh vs. slotted tube balancing radial strength, longitudinal flexibility, and metal-to-artery coverage ratio, tube crimped over angioplasty balloon catheter at a small first (delivery) diameter d
Closest prior art: JP5908937B2 — method of making an endoprosthesis by forming ring struts/hinge elements/connecting struts into defined cell geometry and crimping to a delivery state (a fabrication-plus-crimp process); JP2010246987A also discloses forming an interconnected-strut porous tubular wall. (verify)
Distinguishing limitation: A method of producing and preparing the deformable scaffold defined by the specific geometry selection that balances radial strength, longitudinal flexibility, and metal-to-tissue coverage ratio in a plastically-deformable metal (in particular the wire-mesh construction), then crimping over the balloon — as distinguished from JP5908937B2, which is directed to a polymer-tube molecular-orientation process with a particular W-shaped closed-cell strut architecture. The surviving distinction is the metal wire-mesh geometry and coverage-ratio trade-off, not the generic 'crimp onto balloon' step. Counsel should assess whether the geometry trade-off is described with enough specificity to support a process distinction.
Standard: §103 obviousness (for counsel to assess)
Novelty & Nonobviousness Stress Test
Inventive-Step Stress Test
The adversarial pass: the strongest case AGAINST the moat, so counsel sees the threats before an examiner or opponent does. Challenges rest only on the prior art available to this report; inventor-asserted premises are checked against that art rather than repeated as fact. Engineering work-product — it names §102/§103, it does not conclude.
The single greatest threat to the moat is the 1988 clinical paper (Implantation of balloon-expandable intravascular grafts): it discloses a collapsed stainless steel expandable MESH stent crimped over an angioplasty balloon, delivered through a sheath, and balloon-expanded (plastically) in situ in a vessel. That reference simultaneously hits four of the claimed distinctions — the wire-mesh construction (the very limitation the architecture identifies as the surviving one after JP2010246987A takes the slotted tube), the stainless-steel non-shape-memory metal composition, the crimped small delivery profile, and the operator-set in-situ plastic expansion method. Combined with JP2010246987A (plastic strut-tube balloon expansion to an in-situ point) and the corpus of stainless/CoCr balloon-expandable bare-metal references, the broadest mechanism and the wire-mesh limitation are substantially challenged. The FIRST order of business for counsel is the priority/publication-date contest: if the disclosure's priority antedates the 1988 paper and the retrieved patents (many of which appear to post-date the foundational concept), most of this collapses to §103-only arguments over a much thinner field. The strongest SURVIVING position, if dates cooperate, is not 'balloon-expandable plastic metal tube' (well-populated) but the narrower coupled trade-off: a specific metal wire-mesh cell geometry and metal-to-artery coverage ratio that delivers a quantified radial-strength-per-coverage and radial-strength-per-profile performance across a defined operator-selectable d'/d window — i.e., the distinguishing metrics, not the trivially-satisfied absolute strength, absolute profile, or bare expansion-ratio numbers. The composition claim should be anchored to that mechanism rather than the bare metal list, and the process claim will survive only if the wire-mesh geometry/coverage trade-off is described with real specificity, since JP5908937B2 and CN107072773B already treat cell geometry and coverage as known tunable design axes.
Strongest challenges to the distinguishing limitations
-
Wire-mesh tubular metal body that is plastically (not elastically) deformed in situ, with expanded diameter limited by the lumen wall rather than a device-preset maximum — substantial — counsel must weigh
- The 1988 clinical paper describes a 'collapsed stainless steel expandable mesh stent placed over the balloon of an angioplasty catheter' and advanced through a sheath into the pulmonary artery/target vein, then balloon-expanded. That is a wire-mesh (mesh-wall) balloon-expandable stainless steel scaffold crimped on an angioplasty balloon and plastically deformed in situ — precisely the surviving wire-mesh distinction the architecture rests on, not merely the slotted-tube variant. Bears directly on §102/§103 — for counsel to weigh, subject to verifying this paper's 1988 publication date against the disclosure's priority.
- Rests on: Paper — Implantation of balloon-expandable intravascular grafts by catheterization in pulmonary arteries and systemic veins (1988) (provenance: grounded)
-
Slotted thin-walled tube as the plastically deformable body expanded to a wall-limited diameter — substantial — counsel must weigh
- JP2010246987A discloses an unexpanded stent whose tubular wall of interconnected struts (a porous, strut-defined slotted/cut-tube wall) undergoes plastic deformation to a 'maximum allowable point' on radially outward force. This directly discloses balloon-expandable, plastically deformable metal strut-tube deployment; the bare slotted-tube embodiment is closely approached. The one residual gap is that JP2010246987A frames the limit as a device 'maximum allowable point' rather than a lumen-wall-set diameter. Bears on §102/§103 — for counsel to weigh.
- Rests on: JP2010246987A (verify) (provenance: grounded)
-
Composition: non-shape-memory, work-hardening implant-grade metal (stainless steel, tantalum, titanium) holding diameter purely by plastic deformation — substantial — counsel must weigh
- The 1988 paper expressly uses a stainless steel balloon-expandable mesh stent — a work-hardening, non-shape-memory implant metal that retains diameter by plastic deformation. Additional balloon-expandable stainless-steel/CoCr bare-metal references (Cobalt Chromium or Stainless Steel Balloon-Expandable Bare Metal Stents; SLM CoCr balloon-expandable stent) corroborate the metal-plus-plastic-retention pairing. US6059810A is genuinely distinguishable (shape-memory austenitic set diameter), but the plain stainless-steel plastic-retention pairing is disclosed by the 1988 paper. Bears on §102/§103 — for counsel to weigh.
- Rests on: Paper — Implantation of balloon-expandable intravascular grafts (1988); Cobalt Chromium or Stainless Steel Balloon-Expandable Bare Metal Stents (provenance: grounded)
-
Method of use: plastic deformation in situ to an operator-selected, wall-limited final diameter (variable ratio d'/d chosen at point of use), permanently retained after withdrawal — substantial — counsel must weigh
- The 1988 paper deploys a stainless steel mesh by balloon expansion in the vessel — the final diameter is inherently set by the balloon/vessel wall at the point of use and retained after balloon deflation and withdrawal. This is the operator-set, wall-limited plastic-retention method the claim relies on, not merely 'balloon delivery.' JP2010246987A supplies the multi-stage balloon plastic-expansion sequence. Bears on §102/§103 — for counsel to weigh.
- Rests on: Paper — Implantation of balloon-expandable intravascular grafts (1988); JP2010246987A (verify) (provenance: grounded)
-
Process: metal wire-mesh geometry balancing radial strength, longitudinal flexibility and metal-to-tissue coverage, then crimping over the balloon — substantial — counsel must weigh
- JP5908937B2 teaches the strength/flexibility/cell-geometry trade-off and the crimp-to-delivery step, but in a polymer tube with a specific W-cell architecture. The metal wire-mesh fabrication-plus-crimp is shown by the 1988 paper (stainless steel mesh crimped on balloon). The woven/braided metal-wire mesh constructions of US9925074B2, US20200046527A1, and JP2016116633A show mesh wire geometry generally, though those are self-expanding. The coverage-ratio/cell-configuration trade-off is a recognized design axis (CN107072773B lists strut thickness, cell/aperture shape, cut pattern as tunable parameters). The surviving process distinction depends entirely on whether a specific metal wire-mesh geometry/coverage ratio is described with enough particularity — if only generic, the process claim is heavily approached. Bears on §102/§103 — for counsel to weigh.
- Rests on: JP5908937B2; Paper (1988); US9925074B2; CN107072773B (verify) (provenance: grounded)
Inventor-asserted premises, checked against the art
-
Prior self-expanding intraluminal grafts and springs expanded to a single predetermined diameter the operator could not control (industry_norm) — supported by the provided art
- D1 (self-expanding grafts) and D2 (coil/spring endoprostheses) are cited as fixed-diameter self-expanding devices; the self-expanding woven/braided shape-memory references (US9925074B2, US20200046527A1) confirm the self-expanding category recovers toward a preset shape. US6059810A also relies on a set austenitic recovery diameter. Note the nuance: self-expanding devices are still to some degree constrained by the vessel, so 'wholly operator-uncontrollable' may be overstated — but the core assertion of a factory-set nominal diameter is corroborated.
-
Predetermined-diameter grafts either fail to appose the vessel wall (migration/incomplete scaffolding) or over-expand and risk rupturing the vessel (comparative_advantage) — uncorroborated (inventor-asserted; not shown by the provided art)
- No provided reference quantifies or documents the specific dual failure mode (mal-apposition/migration vs. rupture) of fixed-diameter self-expanding grafts. The 'relocatable stent' paper notes conventional stents cannot be repositioned once expanded, and 'Expandable Metal Stents: Principles and Tissue Responses' may bear on this, but neither is shown to establish the asserted failure dichotomy. A clinical/complication study of self-expanding graft migration and vessel rupture would corroborate or refute it.
-
Plastic deformation beyond the elastic limit holds the lumen open with no elastic spring-back after balloon deflation (other) — supported by the provided art
- The FE plasticity papers corroborate the mechanism — 'this plastic deformation induces static stresses which will remain for the lifetime of the device' (Finite element comparison, 2006) and the 2003/2007 papers on plasticity during dilation. Caveat for counsel: those same papers acknowledge some elastic recoil always occurs on balloon deflation, so 'no spring-back' is an idealization; the mechanism of predominantly plastic retention is nonetheless supported.
-
Residual radial strength after plastic expansion is sufficient to resist chronic recoil of the vessel wall (performance_target) — supported by the provided art
- The 1988 in-vivo paper demonstrates stainless steel balloon-expandable mesh stents remaining patent in pulmonary/venous placement, and 'Expandable Metal Stents: Principles and Tissue Responses' plus radial-strength design papers (JP5908937B2 titled for high radial strength) support that balloon-expandable metal scaffolds resist recoil in general. This device's own quantitative radial strength is described_not_demonstrated, so the specific number is not shown — but the class-level proposition is corroborated.
-
The crimped delivery profile (diameter d) is small enough for percutaneous catheterization through the vasculature to the target lesion (performance_target) — supported by the provided art
- The 1988 paper describes a collapsed stainless steel mesh stent crimped over an angioplasty balloon and advanced through a long sheath introduced over a wire into the pulmonary artery/target vein — a demonstrated small percutaneous delivery profile for exactly this construction.
-
The invention provides controlled, in-situ plastic expansion matched to each individual vessel, solving the one-size problem of prior self-expanding grafts (comparative_advantage) — contradicted by the provided art
- The 1988 paper already achieves operator-controlled in-situ balloon plastic expansion of a stainless steel mesh stent, and JP2010246987A discloses balloon plastic expansion to an in-situ point — so the asserted advantage over self-expanding grafts, while real relative to D1/D2, is not unique to this invention; the prior balloon-expandable mesh art delivers the same 'variable, in-situ, wall-set' expansion. Counsel must confirm the relative priority dates.
Performance-metric scrutiny
-
Residual radial strength sufficient to resist chronic vessel recoil — trivially satisfied by a conventional element — not where the novelty lives
- Why: Any stiff, thick-walled or dense-coverage metal tube delivers large absolute radial strength for free; a solid work-hardened metal scaffold makes 'holds the lumen open' easy to satisfy.
- Metric that actually distinguishes the invention: Radial strength achieved at low metal-to-artery coverage ratio and low crimped crossing profile — i.e., radial-strength-per-coverage (or per-profile), the actual wire-mesh vs. slotted-tube trade-off.
-
Crimped delivery profile (diameter d) small enough for percutaneous access — trivially satisfied by a conventional element — not where the novelty lives
- Why: A thin, sparse wire mesh gives a small crimped profile trivially; minimizing profile alone is easy if radial strength is disregarded.
- Metric that actually distinguishes the invention: Small crimped profile WHILE retaining post-expansion radial strength and adequate coverage — the coupled profile-vs-strength trade-off, not profile alone.
-
Variable, operator-selected expansion ratio d'/d matched to the individual vessel — trivially satisfied by a conventional element — not where the novelty lives
- Why: A large nominal d'/d is trivially obtained by starting from a very small crimped diameter; any balloon-expandable open mesh can be over-expanded to a large ratio in the abstract.
- Metric that actually distinguishes the invention: Retained radial strength and stable wall apposition maintained across the full range of achievable d'/d (i.e., performance sustained over the operator-selectable diameter window), not the magnitude of the ratio.
Claim Outline
Claim Scaffold
A drafting scaffold for counsel — plain-language claim CONCEPTS and fallback positions to turn into claims, NOT drafted claims and NOT legal advice. The broadest claim is on the mechanism (it covers every use); each dependent rung is a narrower fallback if the independent claim is narrowed. Every item is tagged by the disclosure support behind it (§112).
Independent claim — Structure
Concept: A radially deformable open tubular metal scaffold whose wall is an interlaced wire-mesh construction, delivered in a compact crimped state on an internal dilating member and permanently enlarged in place by plastic deformation past the metal's elastic limit, such that after the dilating force is removed the mesh tube holds an expanded diameter that is limited and set by the surrounding lumen wall rather than by any diameter fixed during manufacture. The claim should rest on the mesh-wall body together with the wall-limited (not device-limited) retained diameter, not on plastic deformation to a retained diameter standing alone. (§112 support: described, no data)
Core elements: tubular prosthesis formed as a wire-mesh tube, radial plastic deformation with force in excess of the elastic limit of the material, tube retains expanded diameter d' after balloon deflation and catheter withdrawal, final implanted diameter d' set in situ by the vessel wall, not predetermined at manufacture
Dependent ladder (broad → narrow):
- the mesh formed of an implant-grade, work-hardening (non-shape-memory) biocompatible metal such as stainless steel, tantalum, or titanium — separates the retained diameter from any phase- or temperature-driven set shape (contrast US6059810A austenitic recovery); commercially the primary metallic stent material set (§112 support: described, no data)
- the wire-mesh wall characterized by a specified metal-to-artery coverage ratio balancing radial support against open cell area — gives the mesh geometry a quantifiable structural signature distinct from a bare deformable tube; covers commercial coverage-ratio tuning (§112 support: described, no data)
- expansion ratio d'/d of the mesh sufficient to scaffold typical arterial lumens on the order of a few millimetres from the crimped delivery diameter — ties the deformability to real vessel-sizing performance; blocks a competitor arguing the mesh cannot reach clinically useful diameters (§112 support: described, no data)
- alternatively the deformable body formed as a slotted thin-walled tube with a defined cell/strut configuration and coverage ratio rather than as bare slotted geometry — the bare slotted tube is closely approached by JP2010246987A and JP5908937B2, so this rung claims the slotted embodiment only with narrower cell geometry (§112 support: described, no data)
Independent claim — Composition
Concept: A tubular vascular scaffold body made of a non-shape-memory, work-hardening implant-grade metal that holds an enlarged diameter purely through plastic deformation of the metal, with no temperature- or phase-driven recovery diameter, the retained diameter being limited by the surrounding lumen wall. The durable content is the pairing of the metal with the wall-limited plastic-retention mechanism, not the metal list by itself. (§112 support: described, no data)
Core elements: biocompatible metal candidates: implant-grade stainless steel, tantalum, silver, gold, titanium, radial plastic deformation with force in excess of the elastic limit of the material, final implanted diameter d' set in situ by the vessel wall, not predetermined at manufacture
Dependent ladder (broad → narrow):
- the metal being implant-grade stainless steel — the workhorse balloon-expandable stent material; core commercial embodiment (§112 support: described, no data)
- the metal being tantalum — radiopaque high-ductility alternative; covers a common design substitution (§112 support: described, no data)
- the metal being titanium, gold, or silver — captures the remaining named biocompatible substitutes so the genus of work-hardening metals is not easily circumvented by material swap (§112 support: described, no data)
- the metal selected such that the expanded diameter is retained solely by plastic set with no austenitic/phase-transformation contribution to the final diameter — sharpens the contrast with shape-memory recovery devices (US6059810A) at the composition level (§112 support: described, no data)
Independent claim — Method-of-use
Concept: A method of supporting a body lumen in which a crimped metal tube on an angioplasty balloon catheter at a small delivery diameter d is advanced into the lumen and then plastically deformed in place, by inflating the balloon with force beyond the metal's elastic limit, to an operator-selected larger diameter d' that is limited by the surrounding lumen wall; after balloon deflation and catheter withdrawal the tube permanently retains d', with the expansion ratio d'/d chosen at the point of use to match the individual passageway. The distinction should be framed on the wall-limited, operator-set final diameter and permanent plastic retention, not on balloon delivery generally. (§112 support: described, no data)
Core elements: tube crimped over angioplasty balloon catheter at a small first (delivery) diameter d, radial plastic deformation with force in excess of the elastic limit of the material, tube retains expanded diameter d' after balloon deflation and catheter withdrawal, operator-controlled, variable expansion ratio d'/d matched to individual body passageway
Dependent ladder (broad → narrow):
- the operator selecting d' in situ from a range of achievable diameters so a single delivered device conforms to differently sized lumens — distinguishes factory-preset self-expanding devices (D1/D2) that reach one predetermined diameter (§112 support: described, no data)
- the scaffold remaining as a permanent supporting structure after the vessel is dilated, rather than the vessel being dilated without leaving a retained scaffold — distinguishes plain angioplasty (D3) that dilates without a permanently deformed support (§112 support: described, no data)
- performing the method in a blood vessel to scaffold an arterial lumen of a few millimetres — core clinical use; ties expansion ratio to real vessel sizes (§112 support: described, no data)
- performing the method in a non-vascular passageway such as a biliary, urethral, or esophageal lumen — extends method-of-use coverage to other lumens a competitor might target (§112 support: speculative)
Independent claim — Process
Concept: A method of producing and preparing the deformable scaffold in which a plastically deformable metal body is formed with a geometry selected to balance radial strength, longitudinal flexibility, and metal-to-tissue coverage ratio — in particular an interlaced wire-mesh construction — and is then crimped over a balloon catheter to a small delivery diameter. The surviving distinction is the metal wire-mesh geometry and coverage-ratio trade-off, not the generic crimp-onto-balloon step. (§112 support: described, no data)
Core elements: tubular prosthesis formed as a wire-mesh tube, structural geometry trade-off: wire mesh vs. slotted tube balancing radial strength, longitudinal flexibility, and metal-to-artery coverage ratio, tube crimped over angioplasty balloon catheter at a small first (delivery) diameter d
Dependent ladder (broad → narrow):
- forming the body specifically as an interlaced wire-mesh tube of work-hardening implant-grade metal — distinguishes a polymer molecular-orientation process (JP5908937B2); anchors the process in the mesh metal embodiment (§112 support: described, no data)
- selecting the coverage ratio to a target value trading radial support against flexibility for a chosen vessel size — gives the process a concrete design parameter distinct from generic tube fabrication (§112 support: described, no data)
- alternatively forming the body as a slotted thin-walled metal tube with a defined cell configuration, then crimping — covers the slotted-tube manufacturing route with narrower geometry to stand apart from prior slotted/strut disclosures (§112 support: described, no data)
Blocking claims (for obvious design-arounds)
-
Workaround: Using a shape-memory or superelastic alloy (e.g., nitinol) that self-expands to a factory-set diameter after deployment Block with: Claim coverage tied to the wall-limited plastic-retention mechanism and non-shape-memory work-hardening metal genus, so the operative distinction is that the final diameter is set by the lumen at the point of use rather than by any built-in recovery shape — capturing the plastic-set concept regardless of specific metal (§112 support: described, no data)
-
Workaround: Substituting the wire-mesh wall with a laser-cut slotted or strut-defined tube to avoid a mesh-specific structure claim Block with: A genus rung covering any open-cell plastically deformable metal tubular wall — mesh or slotted — characterized by a specified coverage-ratio/cell-geometry trade-off, so an obvious wall-form swap still reads on the coverage-ratio limitation (§112 support: described, no data)
-
Workaround: Swapping the named metal for another biocompatible work-hardening metal or alloy not on the list Block with: Claim the functional genus of non-shape-memory, plastically deformable implant-grade metal that retains a wall-limited diameter, rather than only the enumerated metals, so a material substitution outside the list still reads on the functional recitation (§112 support: described, no data)
-
Workaround: Delivering with a mechanical/self-expanding dilator or a non-balloon expander instead of an angioplasty balloon Block with: Frame the delivery limitation on any internal radial dilating member applying force beyond the elastic limit, not on a balloon specifically, so an alternate internal dilator still reads on the expansion step (§112 support: described, no data)
-
Workaround: Deploying the same plastic scaffold in a non-vascular lumen (biliary, urethral, esophageal) to escape vessel-specific claims Block with: A method-of-use rung reciting any body passageway with an operator-selected, wall-limited plastic retention, capturing non-vascular applications (§112 support: speculative)
Notes: All embodiments in the disclosure are described but not demonstrated — no working examples or measured parameters (expansion force, achieved d'/d, coverage-ratio values, radial strength data) are provided; nothing here is tagged data_backed, and none should be upgraded without added data. Before filing, counsel should consider whether to obtain: (1) mechanical test data showing plastic set and retained diameter under physiological loading for at least the stainless steel and tantalum mesh embodiments; (2) quantified coverage-ratio and expansion-ratio ranges to support the geometry-trade-off dependent rungs and the process claim, since §112 support for a 'geometry selection' process distinction depends on how specifically the trade-off is described; (3) at least one concrete numerical d and d' pairing to anchor the 'few millimetres arterial' rung. The non-vascular use rung is a reasoned extension only and is tagged speculative. The silver/gold rungs are named but thinly supported — counsel to assess §112 enablement/utility for those metals. Distinctions over JP2010246987A, JP5908937B2, US6059810A, and D1/D2/D3 should be carried on the wall-limited plastic-retention mechanism and mesh/coverage-ratio geometry rather than on plastic deformation or balloon delivery alone, as flagged in the architecture. Novelty/non-obviousness and §112 sufficiency are for counsel to assess.
Prior-Art Differentiation Notes
Per-Application Prior-Art Notes
This is engineering work-product, not a legal opinion or a clearance/validity search. It names the legal standards; the patentability conclusions are the attorney’s.
How to read this section (the §103 framework, stated once): under Graham/KSR, obviousness turns on the scope and content of the prior art, the differences from the claims, the level of ordinary skill, and objective indicia — and any reason to combine references must be articulated, never hindsight. The per-application notes below POSITION each use against that framework; the deeper, claim-level inventive-step analysis is treated separately, at the level of the mechanism’s claims.
Each application is triaged as either "spec support only" (a predictable same-mechanism use — valuable as disclosure / spec breadth, but unlikely to support its own non-obvious claim) or "claim candidate" (required non-obvious adaptation — may merit a claim of its own).
Endovascular / interventional cardiology (origin field) — Balloon-expandable scaffold for maintaining patency of coronary, iliac, femoral, renal and other blood vessels after angioplasty, with final diameter set operator-side against each vessel wall.
Closest existing work: Inventor-cited self-expanding intraluminal grafts (D1) and coil/spring endoprostheses (D2), which self-expand to a single predetermined diameter fixed at manufacture, combined with balloon angioplasty delivery (D3). Among retrieved patents, US6059810A (balloon-mounted shape-memory-alloy stent) and JP2010246987A (balloon-expandable plastically-deformed small-vessel stent) are the closest structurally — but both likely postdate this invention; counsel should confirm priority dates.
Differentiation: Against D1/D2 (self-expanding, fixed predetermined diameter), the named novel elements — radial plastic deformation beyond the material's elastic limit via an internal balloon, with the final implanted diameter d' set in situ by the vessel wall at an operator-controlled variable ratio d'/d rather than preset at manufacture — are the distinguishing features; against retrieved US6059810A the difference is that expansion is by pure plastic deformation of a biocompatible metal (steel/tantalum/etc.) rather than a martensitic-to-austenitic shape-memory phase change. Comparison to D1–D3 is abstract-level (titles/snippets only); US6059810A/JP2010246987A comparison is claim-level but subject to priority-date verification. §102 novelty / §103 obviousness for counsel to assess.
Triage: claim candidate — the application required non-obvious adaptation; it may merit a claim of its own
Citations:
- D1 — Inventor-cited key problem reference — self-expanding grafts limited to a single predetermined diameter, the precise limitation the operator-set in-situ diameter overcomes (provenance: grounded)
- D2 — Inventor-cited category — coil/spring self-expanding endoprostheses expanding to a fixed diameter (provenance: grounded)
- D3 — Inventor-cited delivery/expansion platform (percutaneous balloon dilation) upon which the plastic-deformation mechanism relies (provenance: grounded)
- US6059810A (verify) — Balloon-mounted expandable stent, but expands via shape-memory (Nitinol) martensitic/austenitic phase, not pure plastic deformation of a biocompatible metal; likely postdates — verify priority (provenance: verify_full_doc)
- JP2010246987A (verify) — Balloon-expandable stent undergoing radial plastic deformation to a maximum allowable point; overlaps the plastic-deformation element but likely postdates — verify priority (provenance: verify_full_doc)
Standard: §102 novelty (for counsel to assess)
Inventive-step consideration (§103 argument): Against the inventor's cited art, the §103 inquiry centers on whether a PHOSITA would bridge the plastically-deformable-tube gap: D3 supplies the balloon-catheter dilation platform and D1/D2 supply the concept of an implanted intraluminal scaffold, but both D1 and D2 achieve their final diameter by self-expansion to a single predetermined size, whereas the named novelty is a tube crimped at diameter d and plastically deformed past its elastic limit to an operator-set final diameter d' fixed in situ by the vessel wall. Note that the retrieved balloon-expandable references (US6059810A, JP2010246987A) and the FE-simulation papers appear to postdate a foundational balloon-expandable stent concept and may be anachronistic as prior art — counsel should verify each reference's effective date against the invention's priority date before treating them as §103 art.
Articulated reason (KSR): A PHOSITA seeking to overcome the fixed-diameter mismatch and migration/over-expansion problem D1/D2 present would have a concrete motivation to mount a deformable scaffold on the D3 angioplasty balloon already used to dilate the same lesion, using the same inflation step to expand the scaffold — a predictable combination of the two cited platforms; however, the specific teaching to select a metal expanded beyond its elastic limit so it takes a permanent, non-recoiling, wall-determined set is not supplied by self-expanding D1/D2 (which rely on stored elastic energy to a preset size), so counsel should assess whether the plastic-deformation/variable-d'/d limitation is taught or only reachable through hindsight.
Reasonable expectation of success: A PHOSITA combining a crimped metal tube with balloon dilation would have reasonable expectation that inflation expands and lodges the tube (mechanically straightforward given D3), but would have less assured expectation that a plastically deformed metal mesh/slotted tube would hold d' against chronic recoil without spring-back or unacceptable vessel damage — the very uncertainty the later FE-fracture and dilation-stress papers document — so the expectation as to the permanent wall-set-diameter result is arguably qualified rather than routine.
Secondary considerations (each needs a nexus):
- Long-felt need — nexus: Directly tied to the plastic-deformation/wall-set-diameter feature: D1/D2's fixed predetermined diameter left an unmet need for a scaffold matched to each individual vessel that resists recoil after angioplasty; strong nexus if D1/D2's shortcomings are corroborated by their texts.
- Teaching away — nexus: The asserted failure mode of self-expanding/fixed-diameter grafts (migration, over-expansion) points away from preset-diameter designs toward the operator-set d'; this rests partly on the inventor's characterization of D1/D2 — corroborated only to the extent D1/D2 are actually cited as fixed-diameter problem references, otherwise an uncorroborated asserted norm that a reference showing acceptable self-expanding performance would defeat.
- Commercial success — nexus: Balloon-expandable stents achieved enormous commercial adoption; nexus to the claimed plastic-deformation/variable-expansion features is plausible but must be shown to flow from these specific limitations rather than from angioplasty adoption generally — nexus not yet established on this record.
- Unexpected results — nexus: Any showing that a plastically set metal tube unexpectedly maintained patency without recoil or migration would have nexus to the beyond-elastic-limit / wall-set-diameter elements; on the present record all embodiments are described_not_demonstrated, so no unexpected-results evidence is corroborated.
Standard: §103 obviousness (for counsel to assess)
Oil & gas well construction / downhole completions — Expandable tubulars, liners, and sand screens run compact into an irregular open borehole and plastically expanded (by mandrel/cone or internal pressure) to line, seal, and structurally support the wellbore wall at a diameter dictated by the hole.
Closest existing work: Balloon-expandable, plastically-deformed intraluminal scaffolds from the medical field — closest is a slotted/strut tube crimped small and expanded past its elastic limit to a wall-dictated diameter (JP2010246987A; US6059810A). All retrieved and inventor-cited art is vascular/body-lumen; none addresses a downhole wellbore.
Differentiation: The named novel elements — a plastically-deformed wire-mesh or slotted metal tube expanded in situ to a wall-dictated (not manufacture-set) diameter with an operator-controlled expansion ratio — map onto the vascular art, but no provided reference discloses their application to an open borehole: large-diameter tubulars/liners/screens expanded by a mandrel/cone or internal pressure against formation load, with the sealing and radial-strength-vs-deliverability (screen vs. solid liner) trade-off re-engineered for downhole scale and pressures. Comparison is claim-level against JP2010246987A/US6059810A (body-lumen, small-diameter, balloon-only) and abstract-level against the inventor-cited references (D1–D3).
Triage: claim candidate — the application required non-obvious adaptation; it may merit a claim of its own
Citations:
- JP2010246987A (verify) — Claims an unexpanded strut/porous tube plastically deformed to a maximum point by radially outward force to a wall-conforming diameter — the core mechanism, but small-vessel (~3.5mm) medical scale only, no downhole/formation context. (provenance: grounded)
- US6059810A (verify) — Balloon-expandable stent with longitudinal sections and connecting pieces balancing flexibility/strength — parallels the geometry trade-off element, but body-lumen and balloon delivery only. (provenance: grounded)
- D3 — Inventor-cited balloon angioplasty delivery/expansion platform; abstract-level only — the vascular delivery analog, distinct from mandrel/cone or pressure expansion of downhole tubulars. (provenance: grounded)
Standard: §102 novelty / §103 obviousness (for counsel to assess)
Inventive-step consideration (§103 argument): Under §103, the question is whether a PHOSITA in downhole completions would arrive at in-situ plastic expansion of a compact tubular to a hole-dictated diameter given the provided references — all of which (D1–D3 and every retrieved patent/paper) are cardiovascular stent art. The invention's named novel elements (crimp-on-balloon delivery at diameter d, expansion beyond the elastic limit to a retained diameter d', operator-set final diameter, and the mesh-vs-slotted strength/flexibility/coverage trade-off) read on the medical references, but the argument turns on whether a borehole-completions PHOSITA is even in the analogous art of coronary stenting.
Articulated reason (KSR): The shared physical mechanism (plastically deforming a tube past its elastic limit so it props a non-uniform wall at an as-installed diameter) supplies a facial motivation, but no provided reference bridges cardiology to oil & gas, so any reason to transplant stent geometry into a wellbore rests on the abstract mechanism rather than an articulated teaching — inviting a hindsight/non-analogous-art challenge that a PHOSITA sizing multi-inch borehole liners would not consult millimetre-scale coronary stent references.
Reasonable expectation of success: A PHOSITA would have general confidence that metal tubes plastically deform and hold a set diameter, but the provided references give no expectation for the downhole regime — formation pressures, multi-inch diameters, sealing against irregular rock, and mandrel/cone (not balloon) expansion are all absent from the cited stent art, so success in the well environment is not predictably shown by these references.
Secondary considerations (each needs a nexus):
- Commercial success / very high commercial pull in expandable tubulars — nexus: Weak nexus: the asserted commercial demand is in oil & gas, but the claimed novel features (mesh vs. slotted geometry, balloon-crimp delivery, d'/d expansion ratio) are described for arterial scaffolds; success of downhole products would need to be tied to these specific features rather than to independent well-engineering advances (packers, seals, mandrel tooling).
- Teaching away / different field of use — nexus: The claim that cardiology art is remote from downhole completions rests on an uncorroborated asserted separation between fields; a reference showing oil-and-gas expandable-tubular practice already using in-situ plastic expansion would defeat both the non-analogous-art argument and any teaching-away.
- Long-felt need for conforming to irregular boreholes — nexus: No provided reference establishes a documented unmet need in well construction; the need is asserted by the inventor and lacks corroboration tying it to the specific claimed expansion mechanism.
Standard: §103 obviousness (for counsel to assess)
Non-vascular medical lumens (GI, hepatobiliary, urology, pulmonology) — Balloon-expandable scaffolds for biliary ducts, ureters/urethra, esophagus, colon, and tracheobronchial airways where a stricture must be held open at a lumen-matched diameter.
Closest existing work: Retrieved balloon-expandable plastically-deformed stent art — JP2010246987A (unexpanded tubular strut stent expanded by radially outward force into plastic deformation) and US6059810A (balloon-mounted mechanically expanded stent with longitudinal sections/connecting pieces for flexibility) — plus the inventor-cited self-expanding grafts (D1/D2, fixed predetermined diameter) and balloon angioplasty platform (D3). All retrieved references are directed to vascular/coronary lumens, not non-vascular GI/biliary/urologic/airway lumens.
Differentiation: At the abstract level (JP2010246987A/US6059810A provide claim text but only for vascular use): the invention's named elements — plastic radial deformation beyond the elastic limit with an operator-controlled, variable expansion ratio d'/d set in situ by the individual wall, embodied as either wire-mesh or slotted thin-walled tube — are the same mechanism the retrieved art already claims for blood vessels; what differs here is only the target lumen (biliary, ureter, esophagus, colon, airway), which the retrieved references do not address. §102 novelty and §103 obviousness are for counsel to assess.
Triage: spec support only — a predictable same-mechanism use; valuable as disclosure / spec breadth, but unlikely to support its own non-obvious claim
Citations:
- JP2010246987A (verify) — Claims an unexpanded tubular strut stent that undergoes plastic deformation to a maximum allowable point via radially outward force — the same balloon-expandable plastic-deformation mechanism, but framed for small vessels, not non-vascular lumens (provenance: grounded)
- US6059810A (verify) — Balloon-mounted stent mechanically expanded with repeating longitudinal sections/connecting pieces for flexibility — overlaps the geometry/flexibility trade-off element; vascular context (provenance: grounded)
- D1 — Inventor-cited self-expanding grafts fixed to a single predetermined diameter — the problem reference the in-situ operator-set diameter is meant to overcome; abstract-level only (provenance: grounded)
- D3 — Inventor-cited balloon angioplasty catheter as the delivery/expansion platform crimped-tube delivery relies on; abstract-level only (provenance: grounded)
Standard: §102 novelty (for counsel to assess)
Inventive-step consideration (§103 argument): Under §103, the analysis weighs whether extending the disclosed balloon-expandable, plastically-deforming tube (D3's balloon platform carrying a metal tube expanded past its elastic limit to defeat the fixed-diameter limitation of self-expanding grafts D1/D2) to non-vascular lumens is merely a predictable use of the same mechanism in an analogous environment — which under KSR presumptively cuts against non-obviousness for this application. The named novel elements (wire-mesh/slotted geometry, biocompatible metals, in-situ operator-set d'/d) are not re-engineered for non-vascular use; the disclosure itself treats 'other body passageways' as speculative.
Articulated reason (KSR): A PHOSITA facing strictures in soft-walled, patient-variable non-vascular ducts (biliary, ureteral, esophageal, colonic, airway) would recognize the identical problem the invention solves in vessels — a self-expanding graft's single predetermined diameter cannot conform to an individual lumen — and would apply the operator-controlled plastic-expansion mechanism as a finite, identified, predictable solution; this is a concrete articulated reason grounded in shared lumen mechanics, not hindsight.
Reasonable expectation of success: A PHOSITA would likely have a reasonable expectation of success because the expansion physics (crimp low-profile, balloon-force plastic deformation beyond elastic limit, retained d') is diameter- and site-agnostic and the same biocompatible metals are already implant-grade; however, site-specific unknowns (bile/urine corrosion, peristaltic and respiratory wall dynamics, larger required diameters and coverage-ratio trade-offs) introduce some uncertainty that counsel should probe per lumen.
Secondary considerations (each needs a nexus):
- Long-felt need — nexus: Plausible clinical pull for non-collapsing, lumen-matched non-vascular stents, but nexus must tie to the specific named elements (in-situ operator-set d'/d, plastic-deformation retention) rather than to stenting generally; not yet established on this record.
- Unexpected results — nexus: No nexus established — all non-vascular embodiments are described_not_demonstrated or speculative, so no unexpected performance in GI/biliary/urologic/pulmonary lumens is shown.
- Teaching away — nexus: None shown; the retrieved art is overwhelmingly vascular/coronary, which is mere absence of non-vascular disclosure, not a teaching away. Any asserted norm that non-vascular lumens cannot be plastically scaffolded rests on an uncorroborated asserted norm and would be defeated by a reference showing non-vascular expandable stenting.
Standard: §103 obviousness (for counsel to assess)
Downhole remediation / pipeline integrity (subsurface) — Expandable metal patch or sleeve deployed compact across a corroded, split, or leaking casing/tubing section and plastically expanded to seal the defect and restore load-bearing wall support (zonal isolation, casing patch).
Closest existing work: Balloon-expandable, plastically-deformed intraluminal stent art — notably JP2010246987A (an unexpanded stent plastically deformed to a wall-set point by radially outward force) and US6059810A (balloon-mounted stent expanded by mechanical dilation) — which disclose the same core mechanism: a crimped tube plastically expanded in situ beyond the elastic limit so the surrounding wall sets the final diameter. No downhole/pipeline-casing prior art was provided.
Differentiation: Claim-level against the mechanism: the named elements — radial plastic deformation in excess of the elastic limit, retained expanded diameter after tool withdrawal, and operator-set wall-determined diameter — are already recited in the medical stent claims (e.g., JP2010246987A's 'plastic deformation to a maximum allowable point' by radial force). At the domain level (abstract, not claim-level, since no downhole reference exists here) none of the provided references disclose a metal patch/sleeve deployed across a corroded casing/tubing defect for zonal isolation or pressure sealing, but the underlying deformation mechanism is common to the medical art.
Triage: spec support only — a predictable same-mechanism use; valuable as disclosure / spec breadth, but unlikely to support its own non-obvious claim
Citations:
- JP2010246987A (verify) — Discloses an unexpanded tube plastically deformed to a wall-set maximum point by a radially outward force — the exact core plastic-expansion / no-spring-back mechanism the casing patch relies on. (provenance: grounded)
- US6059810A (verify) — Balloon-mounted stent mechanically expanded by an internally applied dilating force; same delivery/expansion platform as the claimed patch (differs by shape-memory alloy phase, not by the plastic-expansion concept). (provenance: grounded)
- CN107072773B (verify) — Multi-segment expandable prosthesis with differing diameters anchored in a lumen and overlapping sleeve segments — abstract-level analogue to a casing sleeve/overlap-seal geometry; verify full text. (provenance: verify_full_doc)
Standard: §102 novelty / §103 obviousness (for counsel to assess)
Condensed entry: triaged spec-support-only, so the full §103 positioning is reserved for claim-candidate applications and the origin field.
Mining / tunneling geotechnical support (rock reinforcement) — Expandable friction rock bolt / ground anchor: a folded thin-walled metal tube inserted into a drilled borehole and inflated to plastically conform to the irregular rock wall along its length, providing frictional anchoring at a hole-set diameter.
Closest existing work: All provided references are medical balloon-expandable (or self-expanding) intraluminal stents — e.g., US6059810A (balloon-mounted stent plastically expanded in a vessel), JP2010246987A (unexpanded stent plastically deformed to a wall-set diameter by radial force), and inventor-cited D1–D3. None is in the mining/geotechnical domain; the disclosure itself flags a known commercial expansion-in-borehole rock-bolt analog that counsel must locate, as it is not in the provided list.
Differentiation: At an abstract level (mining prior art was not provided), the named novel elements — a wire-mesh or slotted thin-walled biocompatible-metal tube crimped to delivery diameter d and plastically deformed beyond its elastic limit to an in-situ, wall-set diameter d' — map one-for-one onto an expandable friction rock bolt, the differences being scale, non-biocompatible/structural steel, and borehole (not vascular) wall; the retrieved medical references disclose the same plastic-expansion-to-a-wall-set-diameter mechanism but not the geotechnical application, while the field's own expandable-bolt analog (not in this list) may already disclose the raw function — §102/§103 for counsel to assess.
Triage: spec support only — a predictable same-mechanism use; valuable as disclosure / spec breadth, but unlikely to support its own non-obvious claim
Citations:
- US6059810A (verify) — Balloon-expandable tubular prosthesis mechanically expanded (plastically deformed) against a wall — same core mechanism, medical domain only (provenance: grounded)
- JP2010246987A (verify) — Unexpanded tube subjected to plastic deformation to a wall-limited diameter by radially outward force — closest to the in-situ, wall-set d' element (provenance: grounded)
Standard: §102 novelty / §103 obviousness (for counsel to assess)
Condensed entry: triaged spec-support-only, so the full §103 positioning is reserved for claim-candidate applications and the origin field.
Civil water/wastewater infrastructure (trenchless rehabilitation) — Internal expandable metal liner or sleeve pulled into a corroded/leaking buried pipe in a reduced-diameter state and plastically expanded to line and structurally reinforce the host pipe at its actual (variable) internal diameter.
Closest existing work: The retrieved art is all medical/vascular: balloon-expandable stents that are crimped at a delivery diameter and plastically expanded in situ (e.g., US6059810A, JP2010246987A) and the inventor-cited balloon angioplasty platform (D3). No provided reference addresses trenchless rehabilitation of buried water/wastewater pipe.
Differentiation: This application recites the SAME named novel mechanism — a metal tube crimped at delivery diameter d, plastically deformed beyond its elastic limit by an internal dilating force, retaining an in-situ variable diameter d' set by the host wall — but applied to a corroded buried pipe rather than a vessel; the closest reference (JP2010246987A) discloses the plastic-deformation-to-variable-diameter element but only for ~3.5 mm arterial lumens, with no disclosure of pipe-lining, pull-in delivery, or infrastructure-scale diameters. Comparison is claim-level against JP2010246987A/US6059810A but the domain gap (pipe vs. vessel) is not addressed by any provided reference — §102/§103 for counsel to assess.
Triage: spec support only — a predictable same-mechanism use; valuable as disclosure / spec breadth, but unlikely to support its own non-obvious claim
Citations:
- JP2010246987A (verify) — Claims an unexpanded stent undergoing plastic deformation to a maximum allowable point via radially outward applied force — the core in-situ plastic-expansion-to-variable-diameter element, though scaled to arteries. (provenance: grounded)
- US6059810A (verify) — Balloon-crimped, mechanically (balloon) expanded tubular stent with longitudinal-flexibility sections — maps to the crimp-on-balloon delivery and structural-geometry elements. (provenance: grounded)
- D3 — Inventor-cited balloon angioplasty catheter dilation — the delivery/expansion platform reused here to expand the liner. (provenance: grounded)
Standard: §102 novelty (for counsel to assess); §103 obviousness of field transfer also implicated
Condensed entry: triaged spec-support-only, so the full §103 positioning is reserved for claim-candidate applications and the origin field.
Candidate Filings & Family Structure
Filing Strategy
File a comprehensive 'jumbo' provisional bundling the core mechanism — a compact tube delivered into a lumen and plastically expanded beyond the elastic limit so it holds a wall-set diameter with no spring-back — anchored on the origin (endovascular) embodiments where disclosure is strongest. Draft one shared detailed description that carries the full material set (stainless, tantalum, silver, gold, titanium), both geometries (wire-mesh and slotted), and the quantified expansion ratio, then convert that shared spec into a US non-provisional plus a PCT with differentiated claim sets tuned per field. Because the far-field industrial applications read as distinct products rather than mere embodiments, spin the subsurface/geotechnical/infrastructure cluster into its own provisional so their prosecution and prior-art posture stay independent of the medical family. Order by commercial pull and evidentiary readiness: medical first (data closest), industrial second (needs build-out).
Recommended Filings
-
provisional — Jumbo provisional covering the core plastically-expandable wall-set-diameter mechanism as embodied in vascular AND non-vascular medical lumens (applications 1 and 3), with the full material palette, wire-mesh and slotted geometries, and quantified d'/d expansion ratio in the shared description.
- Locks an early constructive-reduction date across the two near-field medical applications where disclosure is described (not merely speculative). Highest commercial pull and broadest in-field breadth; establishes the master shared spec other conversions inherit. Deadline/priority timing: verify with counsel.
-
provisional — Separate provisional covering the industrial/subsurface cluster — expandable tubulars & sand screens (2), downhole casing patch/remediation (4), mining friction rock bolt/ground anchor (5), and trenchless pipe rehabilitation liners (6) — framing the borehole/host-pipe as a non-uniform lumen whose supported diameter is set in situ.
- These read as distinct products with distinct commercial drivers and a different prior-art landscape (notably an established commercial expansion-in-borehole analog in mining). Keeping them in a separate provisional isolates that prior-art exposure from the medical family and preserves independent field-of-use licensing. Timing: verify with counsel.
-
non_provisional — US non-provisional converted from the medical jumbo provisional's shared description, with a claim set focused on the endovascular embodiment and dependent claims reaching non-vascular body passageways.
- Files the strongest, most demonstrable field first; shares the description so conversion cost is reduced. Whether closest prior art (self-expanding grafts, coil endoprostheses, balloon angioplasty) is distinguishable is counsel's call.
-
pct — PCT filed off the same shared medical description to preserve foreign options (12-month priority window), enabling later 30-month national-phase entries with jurisdiction-tuned claim sets.
- Defers foreign cost while keeping global optionality for high-pull medical markets. Uses the shared spec so it is a conversion, not a fresh drafting effort. Dates: verify with counsel.
-
divisional — Reserved divisional to carry any subject matter carved out by a restriction requirement in the US non-provisional (e.g., if vascular vs. non-vascular scaffolds are deemed distinct inventions).
- Restriction-driven placeholder that inherits the parent date for the elected-out subject matter; only file if a restriction is imposed.
-
continuation — Reserved continuation to pursue additional claim scope against the same shared disclosure (no new matter) as competitive landscape clarifies.
- Keeps claim strategy flexible while inheriting the original priority date; no new matter added.
Invention Split
Verdict: separate_filing. The medical applications (1 vascular, 3 non-vascular) are one family — the disclosure expressly contemplates 'other body passageways,' the delivery profile, materials, and metal-to-wall coverage trade-offs carry over directly, so 3 reads as an embodiment of 1. The subsurface, geotechnical, and civil-infrastructure applications (2, 4, 5, 6) share the same generalized function but present as different PRODUCTS — different scale, materials, deployment tooling (mandrel/cone, internal pressure), regulatory context, and commercial channels — analogous to the firm's prior composite-vs-nanorod split. They also carry a distinct and, in mining, an acknowledged pre-existing commercial analog that is best quarantined from the medical prosecution. Grouping them separately keeps each moat piece independently prosecutable and licensable. (Whether the shared mechanism is non-obvious across fields is counsel's call.)
- Separate filing candidate: Oil & gas expandable tubulars/liners/sand screens (application 2)
- Separate filing candidate: Downhole remediation / casing-patch sleeve (application 4)
- Separate filing candidate: Mining/tunneling expandable friction rock bolt / ground anchor (application 5)
- Separate filing candidate: Civil water/wastewater trenchless expandable liner (application 6)
Disclosure Gaps (per application)
-
Endovascular / interventional cardiology (1) (described_not_demonstrated): Bench and/or in-vivo data demonstrating plastic expansion beyond elastic limit with quantified recoil (chronic outward force / no spring-back), radial strength vs. deliverability for both mesh and slotted geometries, worked expansion-ratio examples across coronary/iliac/femoral/renal calibers, and biocompatibility support for the listed materials. Currently described but not demonstrated.
-
Non-vascular medical lumens (3) (described_not_demonstrated): Lumen-specific working examples for biliary/ureteral/esophageal/colonic/airway calibers, corresponding expansion ratios and radial-support requirements for soft-walled strictures, migration-resistance data, and any lumen-specific material/coating considerations. Presently generic 'other passageways' language.
-
Oil & gas expandable tubulars / completions (2) (speculative): Downhole-scale structural data — collapse/burst and formation-load performance after plastic expansion, perforated/slotted vs. mesh radial-strength trade-off at completion diameters, expansion via mandrel/cone or internal pressure, sealing/zonal-isolation performance, and materials suited to downhole chemistry/temperature. Needs substantial build-out before it can support a filing.
-
Downhole remediation / casing patch (4) (speculative): Pressure-seal and post-withdrawal apposition data across corroded/split host geometries, load-bearing restoration to the damaged wall, elastomeric/metal sealing interfaces, and demonstration that plastic deformation maintains contact after tool retrieval. Speculative — requires demonstrative data.
-
Mining / tunneling rock bolt / ground anchor (5) (speculative): Frictional anchoring / pull-out load data along a plastically-conformed borehole, folded-tube inflation behavior against irregular rock, and clear articulation distinguishing over the acknowledged existing commercial expansion-in-borehole analog (a prior-art assessment counsel must make). Speculative and prior-art sensitive.
-
Civil trenchless pipe rehabilitation (6) (speculative): Structural reinforcement and standoff/fit data for variable host-pipe IDs, hydrostatic sealing performance, corrosion/service-life considerations for buried water/wastewater duty, and large-diameter expansion feasibility. Speculative — needs performance evidence.
Grounding & Search Log
Grounding Summary
Of 6 application(s) analyzed for prior art:
- 6 grounded (claim text was available for 5 of 14 retrieved references; the comparison is abstract-level for the rest)